Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00500-1
Product Name/Description Concerto & Basic Shower Trolleys (used for assisted hygiene care)

Serial Numbers: 3000846870, 3000852015, 3000856697, 3000856697, 3000856697, 3000646121, 3000674706, 3000674706, 3000676859, 3000674706,
3000717577, 3000717577, 3000720597, 3000720597, 3000756366, 3000766008, 3000783964, 3000783964, 3000783964

Manufactured between 7 March 2014 and 20 April 2015

ARTG Number: 199958
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/04/2016
Responsible Entity Huntleigh Healthcare Pty Ltd
Reason/Issue As part of post market surveillance (complaint reporting) monitoring, ArjoHuntleigh received a limited number of complaints where the bolts under the stretcher became loose. There has been one minor injury as a result of a patient falling related to these incidents.

ArjoHuntleigh investigated the issue and concluded the stretcher mounting bolts or stretcher tilting mechanism can become loose which may make the device unstable, providing a potential for risk to the patient or the caregiver. There is a risk the locking mechanism for the stretcher–tilt function may not engage properly, resulting in a patient falling to the floor.
Recall Action Recall for Product Correction
Recall Action Instructions ArjoHuntLeigh is advising users to ensure the trolleys are checked regularly as per the Instructions for Use (IFU).
ArjoHuntLeigh will be performing an inspection on all affected devices to ensure the bolts are securely fastened. This action has been closed-out on 16/02/2017.
Contact Information 08 9314 4102 - ArjoHuntleigh