TGA Recall Reference RC-2016-RN-00388-1
Product Name/Description CELL-DYN Emerald Cleaner. An in vitro diagnostic medical device (IVD)

List number: 09H46-02

Lot numbers: 6853, 6901
Expiry date: 31 July 2017

Lot number: 6953
Expiry date: 30 Sep 2017

Lot Numbers: 6991
Expiry date: 30 Nov 2017

Lot Numbers: 7024, 7027
Expiry date: 31 Dec 2017

Lot Number: 7044
Expiry date: 31 Jan 2018

Lot Number: 7082
Expiry date: 28 Feb 2018

Lot Number: 7110
Expiry date: 31 Mar 2018

Lot Number: 7119
Expiry date:30 Apr 2018

ARTG number: 225068
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/04/2016
Responsible Entity Abbott Australasia Pty Ltd Diagnostic Division
Reason/Issue Abbott has identified occurrences where the CELL-DYN Emerald analyser generates Quality Control (QC) Out Of Range Low for parameters RBC and PLT. Abbott is continuing to investigate but immediate actions by the laboratory are necessary. Abbott has no evidence that there is impact to generated patient results. There is a potential for delay in results due to QC out of range.
Recall Action Recall
Recall Action Instructions Abbott is advising users to switch to an alternative lot of Cleaner reagent if available, and destroy any remaining inventory of the affected lots. If users are not experiencing the error and do not have an alternative lot to use, please order a replacement and continue to use the reagent provided internal QC results remain in specification. Users are advised to contact customer support if they are experiencing the error and do not have an alternative lot available. This action has been closed-out on 26/05/2017.
Contact Information 1800 816 696 - Abbott Customer Support