Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00388-1 |
Product Name/Description |
CELL-DYN Emerald Cleaner. An in vitro diagnostic medical device (IVD)
List number: 09H46-02
Lot numbers: 6853, 6901 Expiry date: 31 July 2017
Lot number: 6953 Expiry date: 30 Sep 2017
Lot Numbers: 6991 Expiry date: 30 Nov 2017
Lot Numbers: 7024, 7027 Expiry date: 31 Dec 2017
Lot Number: 7044 Expiry date: 31 Jan 2018
Lot Number: 7082 Expiry date: 28 Feb 2018
Lot Number: 7110 Expiry date: 31 Mar 2018
Lot Number: 7119 Expiry date:30 Apr 2018
ARTG number: 225068 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
5/04/2016 |
Responsible Entity |
|
Reason/Issue |
Abbott has identified occurrences where the CELL-DYN Emerald analyser generates Quality Control (QC) Out Of Range Low for parameters RBC and PLT. Abbott is continuing to investigate but immediate actions by the laboratory are necessary. Abbott has no evidence that there is impact to generated patient results. There is a potential for delay in results due to QC out of range. |
Recall Action |
Recall |
Recall Action Instructions |
Abbott is advising users to switch to an alternative lot of Cleaner reagent if available, and destroy any remaining inventory of the affected lots. If users are not experiencing the error and do not have an alternative lot to use, please order a replacement and continue to use the reagent provided internal QC results remain in specification. Users are advised to contact customer support if they are experiencing the error and do not have an alternative lot available. This action has been closed-out on 26/05/2017. |
Contact Information |
1800 816 696 - Abbott Customer Support |