Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00368-1
Product Name/Description Dimension Vista 500/ Dimension Vista 1500 Intelligent Lab Systems, software versions V. 3.6.1 SP1 or V. 3.6.2. An in vitro diagnostic medical device (IVD)

Siemens Material Numbers (SMN): 10488224 and 10444802 respectively

ARTG Number: 174699
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 30/03/2016
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue
Siemens Healthcare Diagnostics has received customer complaints of discrepant patient results on Dimension Vista Intelligent Lab Systems. Siemens Healthcare Diagnostics has identified a software defect which, in a very specific set of circumstances, results in the Dimension Vista System omitting an aliquot probe rinse between sample aspirations when processing tubes in Sample Racks that are front loaded on the Dimension Vista System. Omission of the aliquot probe rinse may result in carryover of residual sample estimated up to 10% when the sample is dispensed into the aliquot well. Carryover of residual sample from the outside of the aliquot probe into the sample tube is estimated at less than 0.1%.

Although unlikely, the potential for erroneous results due to sample carryover may lead to result misinterpretation.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens has developed a software update which will address the reported issue. The update will be installed on all affected units. Until the affected software version is updated, users are advised to follow the specific instructions provided under “Actions to be Taken by the Customer” in the customer letter.
This action has been closed-out on 23/03/2018.
Contact Information 1800 310 300 - Siemens Technical Support Centre