Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00346-1
Product Name/Description INTELEPACS Versions 4.5.1 to 4.12.1
(Radiology picture archiving and communication system)

Affected software identifier contains an “R” but with a number below the following:

PACS-4-5-1-R77
PACS-4-6-1-R85
PACS-4-8-1-R46
PACS-4-10-1-R36
PACS-4-11-1-R28
PACS-4-12-1-R4

ARTG Number: 149923
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/04/2016
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Reason/Issue Due to the asynchronous nature of image ingestion processing, a failure at the early stages of the ingestion process may occur leading to images not being visible to the end user, without any indication to the user that this has occurred.

The user may not be aware that images might be missing. A misdiagnosis or a delay in diagnosis and treatment can occur if health professionals base their diagnosis on incomplete image sets.
Recall Action Recall for Product Correction
Recall Action Instructions In order to correct this problem, Intelerad is:

1- providing concerned clients with a list of affected studies and possibly recover the missing images;
2- deploying a software update that includes both a fix and a mitigation for all clients potentially affected by the problem; and
3- periodically auditing all concerned client systems to find and report to the concerned client any new occurrences of this problem until the software update is in place. This action has been closed-out on 27/01/2017.
Contact Information 02 9006 1662 - Emergo Australia