Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00340-1
Product Name/Description Brilliance CT and Ingenuity CT with software versions v4.13 and v4.1.4 (Computed tomography full-body x-ray system)

ARTG Number: 98868
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 23/03/2016
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue On the affected CT systems and software versions, the issue relates to the auto transfer to the Remote devices (such as PACS), where there is a possibility for intermittent failures to send all images/data series automatically. The images remain on the local storage device until manually transferred to the remote device.
Recall Action Recall for Product Correction
Recall Action Instructions Philips is advising users of affected CT systems to assure that all images are transferred by checking that images are available on any remote devices before images/data series are deleted from local storage device. If there are missing images/data series on remote devices, the user can manually transfer images/data series to remote devices. Philips will be completing a software upgrade as a permanent fix. This action has been closed-out on 23/06/2017.
Contact Information 1800 251 400 - Philips Customer Care Centre