Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00242-1
Product Name/Description SIGNA Pioneer MRI System

ARTG Number: 252692
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 3/03/2016
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue There is a potential safety issue with a missing ferrous warning label on the SIGNA Pioneer Table Control Box (TCB).

Ferrous material in the TCB could cause unexpected attraction to the MR system’s high-strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field.

There have been no events reported as a result of this issue. Clinical scanning of patients is NOT affected by this safety notification.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare will correct all affected products at no cost to the affected customers by applying the ferrous warning label to the TCB. These labels are expected to be available and installed by June 2016. A GE Healthcare representative will contact the affected customers to arrange for the correction.
Contact Information 1800 659 465 - GE National Call Centre