TGA Recall Reference RC-2016-RN-00233-1
Product Name/Description Safe-T-Vue 6 and Safe-T-Vue10 Temperature Indicators (used to monitor the temperature of blood bags during transit)

Multiple part numbers and lot numbers affected

ARTG Number: 265484
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/03/2016
Responsible Entity Helena Laboratories Australia Pty Ltd
Reason/Issue The manufacturer, Temptime is making a change to Safe-T-Vue’s instructions for use. The instructions on the box and package insert state that indicators must be removed from refrigeration and brought to room temperature every 3 months. Temptime conducted additional testing and determined that Safe-T-Vue indicators must be removed from refrigeration for 24 to 48 hours every 2 months. Storing Safe-T-Vue in refrigerated conditions for greater than 2 months may lead to failure of the indicator to change colour when the specified temperature has been reached.
Recall Action Recall for Product Correction
Recall Action Instructions Helena Laboratories is providing updated Instructions For Use (IFU) for users to follow.
Contact Information (03) 9543 7299 - Helena Laboratories Australia Pty Ltd