Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00206-1
Product Name/Description Verso Shoulder Forked Retractor (Comprehensive Instrumentation (2 Prong Retractor))

Part number: 402852
All Lots affected

ARTG Number: 177811
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 24/02/2016
Responsible Entity Biomet Australia Pty Ltd
Reason/Issue An investigation has identified that Verso Shoulder Forked Retractor Part Number 402582 was manufactured using 420 stainless instead of 420 S29 stainless steel. 420 stainless steel is more brittle than 420 S29. The instrument may fracture during use.

If the retractor fractures during surgery, surgical intervention may be necessary to retrieve any fractured pieces. A delay in surgery greater than 30 minutes may occur. If the pieces are unable to be retrieved, the patient would retain a foreign body.
Recall Action Recall
Recall Action Instructions The sponsor is advising users to inspect stocks and quarantine the affected units. The sponsor will remove and replace the affected units. This action has been closed-out on 25/01/2017.
Contact Information 02 9483 5426 - Zimmer Biomet