TGA Recall Reference RC-2016-RN-00200-1
Product Name/Description ITD VESA 75/100 Adaptation with Rotation (Articulating Arm on Endoscope Trolley)

Lot Numbers: HA.2200.902 and HA.2237.902
Multiple Serial Numbers affected

Manufactured between 1 June - 26 November 2015.

ARTG Number: 168578
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/02/2016
Responsible Entity Karl Storz Endoscopy Australia Pty Ltd
Reason/Issue A manufacturing defect has been detected which could lead to the loosening of the rivets of the interface plate that connects the monitor and touch screen to the articulating arms on the trolley (VESA 75/ 100 adaptation with rotation). The riveting of the rotation may not have been carried out according to the manufacturer's specifications.
In rare cases, it can lead to a loosening of the connection plate and result in injuries to patients and/or users.
Recall Action Recall for Product Correction
Recall Action Instructions KARL STORZ is advising customers that they can opt for their own engineers to replace the adaptor plates or nominate KARL STORZ to replace the affected component on site.
Contact Information 02 9490 6700 - Karl Storz