Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00188-1
Product Name/Description iMDsoft MetaVision ICU (MV-ICU), Version 5.47.4622
(graphical patient information system used for data collection, management, and storage)

ARTG Number: 165518
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 22/02/2016
Responsible Entity IMDsoft
Reason/Issue A software issue has been identified in the medication management module which may lead to duplication of medication dose under specific settings. This can only occur when a medication order has a weekly interval (e.g., administer medication X every Monday) and an attribute (e.g, dose quantity or frequency) of the order is subsequently modified.
Recall Action Recall for Product Correction
Recall Action Instructions A temporary mitigation was provided by means of modifying the software configuration setting. Specifically, the proposal was to extend the time horizon from 7 days, which is equal to the dose interval, to greater than 7 days. A permanent solution will be provided once it becomes available.