TGA Recall Reference RC-2016-RN-00179-1
Product Name/Description Berichrom Antithrombin III Kits. An in vitro diagnostic medical device (IVD)

Catalogue Number: OWWR15, OWWR17

Siemens Material Number (SMN): 10446672, 10446673

Lot Number: 45029, 45027

ARTG Number: 178559
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 15/02/2016
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare Diagnostics has confirmed that some packages of the Berichrom Antithrombin III kit may contain a different substrate reagent lot as given in the package labelling. Dependent on the instrument used, some instruments with automatic reagent identification may request a new calibration.
The correct substrate reagent lot number is 520294.
This issue has no impact on patient results.
Recall Action Recall
Recall Action Instructions Siemens Healthcare is advising users that the affected product will be replaced with unaffected stock. This action has been closed-out on 03/02/2017.
Contact Information 1800 310 300 - Siemens Technical Support Centre