TGA Recall Reference RC-2016-RN-00174-1
Product Name/Description Rotor Yokes (used in Allegra X-5 Centrifuge). An in vitro diagnostic medical device (IVD)

Rotor Yokes Part Number: B30593

Lot Numbers: 322/13 and 355/13

ARTG Number: 177999
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 11/02/2016
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue A defective rotor yoke allows improper rotor seating on the hub resulting in possible rotor failure if the buckets are loaded in an imbalanced configuration.
If the rotor becomes unbalanced it may vibrate excessively and/or exhibit smoke if the rotor is rubbing inside the centrifuge.

There is a chance of physical injury to users as a result of the vibration if the issue occurs. There is also a remote risk of a delay in results if the failure occurs.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is replacing the affected rotor yokes. The centrifuge can continue to be used, as long as all runs are balanced appropriately. This action has been closed-out on 07/09/2018.