Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00173-1 |
Product Name/Description |
Klikfit CT Syringe with QFT and 75cm Extension tube
Catalogue Number: ZY6022 All lot numbers with an expiry date up to and including November 2020
ARTG Number: 127663 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
11/02/2016 |
Responsible Entity |
|
Reason/Issue |
An issue has been identified with the 75cm extension tube included with the Klikfit CT Syringe with QFT. The coil presentation of the tube causes a pull force that could dislodge or move the connected cannula which can lead to extravasation. |
Recall Action |
Recall |
Recall Action Instructions |
Imaxeon is advising users to inspect stock and quarantine any remaining units. Affected stock can be returned to Imaxeon and replaced with an alternate product. |
Contact Information |
1800 633 723 - Imaxeon Customer Service |