Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00153-1 |
Product Name/Description |
Colleague P1.7 CXE Single Channel Pump and Colleague P1.7 CXE Triple Channel Pump (Infusion Pump)
Colleague P1.7 CXE Single Channel Pump Product Code: 2M91617
Colleague P1.7 CXE Triple Channel Pump Product Code: 2M91637
ARTG Number: 177494 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
9/02/2016 |
Responsible Entity |
|
Reason/Issue |
Baxter Healthcare has identified that the incorrect version of the operator manual was supplied with the delivery of some machines. The manuals are missing the addendum with instructions regarding the user to reopen the regulating clamp (if one exists above the pump) on the primary set after the secondary infusion has completed.
If the administration set of the primary infusion has a closed clamp, it may lead to air embolism or delay of therapy for the primary infusion. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Baxter Healthcare is providing users with the correct user manual. This action has been closed-out on 15/02/2017. |
Contact Information |
02 8845 1669 - Baxter |