Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00151-1 |
Product Name/Description |
COULTER ReticPrep Reagent Kit (used for reticulocyte counting). An in vitro diagnostic medical device (IVD)
Reference Number: 7546978 Lot Number: 108912K Expiry Date: 11/12/2016
ARTG Number: 223049 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
8/02/2016 |
Responsible Entity |
|
Reason/Issue |
Beckman Coulter has received an increased number of customer complaints related to the inability to scan or manually enter the ReticPrep barcode label.
The ReticPrep Barcode Label number 2 contains characters that are not recognised by the software ("+$$" instead of "22"), and due to this issue the system will not generate results.
The reported issue may lead to a delay in test results reporting. However, it is unlikely to cause adverse health consequences. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Beckman Coulter is advising users to follow the IFU to manually enter barcode number 2 with "22" in place of "+$$" which is printed on the label. |
Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |