Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00149-1
Product Name/Description PRN 50-M+ Digital Writer

Part Number: 600-23310-01
Units manufactured from May 28, 2013 through to April 7, 2015
Serial Numbers: 1801405 through 1918761

ARTG Number: 117497
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/02/2016
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue The manufacturer has recently become aware of a potential safety issue associated with the flammability and fluid ingress rating of the outer enclosure material. There is an increased user/patient risk of exposure to fire and/or electrical shock due to the failure of the PRN 50-M+ Digital Writer to meet its intended flammability and fluid ingress rating.
Recall Action Recall for Product Correction
Recall Action Instructions GE Healthcare is replacing all affected printers with unaffected units. This action has been closed-out on 22/02/2017.
Contact Information 1800 659 465 - GE National Call Centre