Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00149-1 |
Product Name/Description |
PRN 50-M+ Digital Writer
Part Number: 600-23310-01 Units manufactured from May 28, 2013 through to April 7, 2015 Serial Numbers: 1801405 through 1918761
ARTG Number: 117497 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
8/02/2016 |
Responsible Entity |
|
Reason/Issue |
The manufacturer has recently become aware of a potential safety issue associated with the flammability and fluid ingress rating of the outer enclosure material. There is an increased user/patient risk of exposure to fire and/or electrical shock due to the failure of the PRN 50-M+ Digital Writer to meet its intended flammability and fluid ingress rating. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
GE Healthcare is replacing all affected printers with unaffected units. This action has been closed-out on 22/02/2017. |
Contact Information |
1800 659 465 - GE National Call Centre |