Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00145-1 |
Product Name/Description |
Surgical Clipper Charger
Product Code: DYND70802AU Lot Numbers: 1, 2, 3 and 4
ARTG Number: 218048 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
4/02/2016 |
Responsible Entity |
|
Reason/Issue |
Medline have received reports that due to the potential defective component on the circuit board, the clipper charger can overheat. This could result in a clipper charger malfunction and may cause thermal injuries to the user. |
Recall Action |
Recall |
Recall Action Instructions |
Medline are advising users to discard the affected chargers. Medline will replace all affected chargers, and is advising users to use an alternative method of hair removal until a replacement arrives. This action has been closed-out on 14/02/2017. |