Type of Product Medical Device
TGA Recall Reference RC-2016-RN-00052-1
Product Name/Description Central Venous Catheter and Pressure Monitoring Sets

Single Lumen Central Venous Catheter Sets & Trays
Femoral Artery Pressure Monitoring Catheter Sets & Trays
Radial Artery Pressure Monitoring Catheter Sets & Trays

Reference Part Numbers (RPN): C-NPMS-501J-15, C-PMS-301-FA, C-PMS-301J-FA, C-PMS-301J-PED, C-PMS-301-RA, C-PMS-400-RA, C-PMS-401-FA and C-PMS-401J-FA

Multiple Lot Numbers

ARTG Numbers: 154878 and 156190
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 14/01/2016
Responsible Entity William A Cook Australia Pty Ltd
Reason/Issue Certain Central Venous Catheters and Pressure Monitoring Sets are being recalled due to an issue being identified with a manufacturing process that could result in the potential for catheter tip fracture and/or separation.

Potential adverse events that may occur as a result of catheter tip fracture and or separation may include loss of device function, medical intervention to retrieve a separated segment, or complications resulting from a separated tip occluding blood flow to end organs. There have been no reports of illness or injury associated with this issue.
Recall Action Recall
Recall Action Instructions The sponsor is advising users to inspect stock and quarantine the affected units prior to their return to the sponsor.

This action has been closed out on 15/11/2018
Contact Information 1800 777 222 - Cook Medical