TGA Recall Reference RC-2016-RN-00024-2
Product Name/Description One Step HCG Urine Pregnancy Test. An in vitro diagnostic medical device (IVD)

Batch Numbers: 20140406 and 20150314

ARTG number: 260088
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 8/01/2016
Responsible Entity Murray International Trading Pty Ltd
Reason/Issue Subsequent to complaints regarding false negative results by the users, results of the testing undertaken by the TGA for the above batches have showed that these batches do not comply with the claim in the Instructions for Use which specify that “In less than 5 minutes, levels of HCG as low as 25mIU/ml can be detected”. Batch # 20150314 has failed to detect hCG in human urine at concentrations of 10, 25, and 50 mIU/mL levels while batch # 20140406 has failed at hCG concentrations of 10, 25, 50 and 100 mIU/mL . The TGA is undertaking further investigation to ascertain if any further actions are required.

False negative results in the worst case can result in misdiagnosis and significant harm.
Recall Action Recall
Recall Action Instructions Customers are advised to quarantine any unused stock from the above batches and return it to Murray International Trading Pty Ltd.