Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00021-2 |
Product Name/Description |
Cook Beacon Tip Angiography Catheters
Beacon Tip Torcon NB Advantage Catheters Catalogue Prefix: HNBR4.0, HNBR4.1 and only HNBR5.0 w/RUC Suffix
Beacon Tip Royal Flush Plus High-Flow Catheters Catalogue Prefix: HNR4.0
Slip-Cath Beacon Tip Catheters Catalogue Prefix: SCBR4.0, SCBR4.1, and only SCBR5.0 w/RUC Suffix
Shuttle Select Slip-Cath, 4 FR Shuttle System Compatible Catalogue Prefix: SCBR4.5
All Lot Numbers affected
ARTG Numbers: 137861, 154909, 154911, 137865 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
6/01/2016 |
Responsible Entity |
|
Reason/Issue |
Cook Medical is initiating a recall of all lots in distribution following reports of catheter tip splits and/or separation. Potential adverse events that may occur as a result of catheter tip splitting and or separation may include the loss of device function, medical intervention to retrieve the separated segment, or complications resulting from the separated tip occluding blood flow to end organs. |
Recall Action |
Recall |
Recall Action Instructions |
Cook Medical is advising users to inspect and quarantine any affected lots for return to Cook Medical.
This action has been closed out on 25/02/2019 |
Contact Information |
1800 777 222 - Cook Medical |