Type of Product |
Medical Device |
TGA Recall Reference |
RC-2016-RN-00014-2 |
Product Name/Description |
ImmunoCAP Specific IgG/IgG4 i1 Control H. An in vitro diagnostic device (IVD)
Part Number: 10-9475-01
Lot Numbers: All lot numbers when used in combination with ImmunoCAP Allergen i1 Honey bee venom art. No 14-4143-01, lot numbers 389B7 and 389B8
ARTG Number: 186054 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
15/01/2016 |
Responsible Entity |
|
Reason/Issue |
The ranges for mean concentration of IgG/IgG4 stated on the vial label of the quality control sample are incorrect. The new measuring ranges for ImmunoCAP Specific IgG/IgG4 i1 Control H, all lots when used in combination with ImmunoCAP i1, lot 389B7 and 389B8 are:
IgG4: 12 – 25 mg/l IgG: 24 - 47 mg/l
The problem can be detected by the operator since the quality control samples will fail. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
The sponsor is advising users to use the following new measuring ranges for ImmunoCAP Specific IgG/IgG4 i1 Control H: IgG4: 12 – 25 mg/l IgG: 24 - 47 mg/l Patient results are not affected. No action needs to be taken in regards to past or current patient results for ImmunoCAP Allergen i1, Honey bee venom. |
Contact Information |
1800 222 287 - Abacus ALS |