Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01259-1
Product Name/Description Coulter S-CAL Calibrator Kits used on the Unicel DxH 800, DxH 600, LH 750 and LH 780 Systems.

Catalogue Numbers: 624519, 179310 and 628026

Multiple Kits and Vial Lot Numbers affected

An in vitro diagnostic device (IVD)

ARTG Number: 223050
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 23/12/2015
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue Beckman Coulter has received an increased number of customer complaints observed after S-CAL calibration on the Unicel DxH 800 and DxH 600 Coulter Cellular Analysis Systems, and the Coulter LH 750 and LH 780 Hematology Systems.

Customers have reported a variety of issues during calibration including:
- An increase in H&H Check messages
- Low MCH recovery in XB
- Low MCHC recovery in XB
These issues are not observed with all analysers and there is no impact to patient results.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is advising users to monitor other quality control performance and adhere to the established maintenance schedule. Beckman Coulter is adjusting the SCAL values for lots starting 4782 and 113160400. This action has been closed-out on 16/05/2017.