Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01256-1
Product Name/Description Oxylog Transport Ventilator

Models: 2000plus, 3000 and 3000 plus

ARTG Number: 169507
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 23/12/2015
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue Some cases have been reported to Draeger in which the "Poti unplugged" error message was generated. In those cases an audible and visual alarm is generated, the breathing system releases pressure and the ventilation function stops operating. The error message is caused by increased electrical contact resistances of the controllers (setting potentiometers). The increased resistance is caused by an oxide layer that has formed on the controller over a lengthy period of time.

No patient injury has been reported due to this issue.
Recall Action Recall for Product Correction
Recall Action Instructions The oxide layer can only form if the controllers are rarely or never moved and according to product monitoring, some users never, or rarely, use the FiO2 controller. As a preventative measure to ensure the resistance taper of the controllers are sufficiently cleaned, Draeger is recommending the following addition to the daily device check:
- Prior to turning on the Oxylog in readiness to conduct the daily device check, the staff member turns each of the controllers (Vt, RR, Pmax and FiO2) to the right and left stop positions (minimum and maximum values) at least ten times each.
- The Oxylog can then be turned on and the onboard device check completed.
Contact Information 03 9244 7248 - Draeger Medical Australia