Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01246-1
Product Name/Description Samarys Peek Anterior Cervical Cage

Units from the following batches of Samarys Peek Anterior Cervical Cage are affected.
Part No: 11CC2A15-5S Lot No:23LJ-19196
Part No: 11CC2A17-6S Lot No: 23KY+19029
Part No: 11CC2AC17-5S Lot No: 247P+19029
Part No: 11CC2AC17-6S Lot No: 248K+19029

ARTG Number: 101789
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 21/12/2015
Responsible Entity B Braun Australia Pty Ltd
Reason/Issue This Hazard Alert is being issued because of a random non-conformity relevant to the gold wires used as a position marker for the cervical PEEK cage Samarys. The observed facts were that some devices within 4 lots manufactured between April 2014 and November 2015 presented missing, improperly secured, and/or incorrectly positioned gold wire markers, which may protrude 1-2 mm inside or outside the cage.

Gold wires protruding from cage may cause local inflammation and/or neural or vascular injuries at treated site or distant ones by migration (in complete protrusion), which potentially can have serious consequences. The probability of these potential adverse events is however very low. To date, no customer complaint has been recorded and no adverse event or incident has been reported.
Recall Action Hazard Alert
Recall Action Instructions The sponsor is advising the surgeons that in the case of improper fixation of the gold wire, migration of these small dimensioned gold wires in the surrounding tissues would be easily detected on the X-ray films that are routinely taken during patient follow-up. Additionally, as exposed subjects may be often asymptomatic for a long period, surgeons must check on the routine X-ray films that position markers did not migrate. They may also warn implanted patients to report any late dysphagia, sensation of foreign bodies into their throat or suspect cough or different concerning symptoms, as this might allow early detecting a potential complication and solving it. For further information, please see http://www.tga.gov.au/alert/samarys-polyethylethylketone-anterior-cervical-cage This action has been closed-out on 30/08/2016.
Contact Information 02 9629 0200 - B. Braun Australia