Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01238-1
Product Name/Description Dengue NS1 Ag Strip

Catalogue Number: 70700

Serial/Lot Numbers:
5F0130 Expiry Date: 30 March 2016
5H0131 Expiry Dates: 30 April 2016 and 15 May 2016
5H1131 Expiry Dates: 30 April 2016 and 15 May 2016
5H2131 Expiry Dates: 30 April 2016 and 15 May 2016

ARTG Number: 207118

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 22/12/2015
Responsible Entity Bio-Rad Laboratories Pty Ltd
Reason/Issue There has been a sudden degradation of the sensitivity of certain lots of Dengue NS1 Ag Strips, which has resulted in false negative results for positive patient samples.
Recall Action Recall
Recall Action Instructions Bio-Rad is advising users to inspect stock (including units currently in use) and discard the affected units. Affected lots will be replaced. This action has been closed-out on 30/08/2016.
Contact Information 1800 224 354 - Bio-Rad