Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01190-1
Product Name/Description Switch Point Infinity 3 (Operating Room Management System)

Catalogue Number: 0678001000

All serial numbers affected

ARTG Number: 232446
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 7/12/2015
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has become aware that for sources utilising S-video and Composite inputs, Phase Alternating Line (PAL) signals appear with jagged edges on all diagonal lines resulting in poor video quality. The potential hazard associated with this failure is a PAL signal used with S-video or composite input that produces an image that is not clear and free of artefacts.
Recall Action Recall for Product Correction
Recall Action Instructions A Stryker Technical Service Representative will coordinate the update of affected devices. The update will correct the poor video quality as well as support for additional PAL resolution. This action has been closed-out on 29/08/2016.
Contact Information 02 9467 1167 - Stryker