Product Name/Description |
A.L.P.S. 4.0mm Cancellous Locking Screws (non-sterile)
CNCLS LOCK SCREW FT 4.0x10mm, 4.0x12mm, 4.0x14mm, 4.0x16mm, 4.0x18mm, 4.0x20mm, 4.0x22mm ,4.0x24mm, 4.0x26mm, 4.0x28mm, 4.0x30mm, 4.0x32mm, 4.0x34mm, 4.0x36mm, 4.0x38mm, 4.0x40mm, 4.0x42mm, 4.0x44mm, 4.0x46mm, 4.0x48mm, 4.0x50mm, 4.0x55mm, 4.0x60mm, 4.0x65mm, 4.0x70mm, 4.0x75mm, 4.0x80mm, 4.0x85mm, 4.0x90mm, 4.0x95mm, 4.0x100mm
Multiple Part Numbers Affected
All Lot Numbers Affected
ARTG Number: 209305 |
Reason/Issue |
It has been identified that all lots of A.L.P.S. 4.0mm Cancellous Locking Screws, manufactured between June 1, 2008 and June 4, 2015, have a single lead thread instead of a triple lead thread. A single lead screw may sit proud in the plate which could lead to soft tissue irritation. In addition, if the screw does not lock into the plate, it is possible for the screw to back out. In the event that the screw does not lock into the plate and backs out, a revision surgery may be necessary. |