Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01097-1
Product Name/Description Oxoid Antimicrobial Susceptibility Testing Disc Cefoxitin. An in vitro diagnostic medical device (IVD)

Product Code: FOX30 CT0119B

Lot Numbers: 1689092 & 1689093093

ARTG Number: 208273
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 10/11/2015
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue An internal technical investigation by Oxoid Limited, part of Thermo Fisher Scientific, has confirmed that certain lots of Oxoid Antimicrobial Susceptibility Testing Disc Cefoxitin may contain discs that have insufficient concentration of antibiotic. It has been identified that the low concentration may impact on the performance of the disc against Escherichia coli and Neisseria gonorrhoeae quality control organisms resulting in lower than expected zones of inhibition.

Performance with the Staphylococcus aureus QC organism is currently within specification.
Continued use of this lot could result in quality control failures, delayed results or incorrect results reporting.
Recall Action Recall
Recall Action Instructions End users are requested to inspect their stock for affected units. Affected units are to be removed from further use. The requirement for the review of patient results should be determined by the laboratory director. This action has been closed-out on 22/08/2016.
Contact Information 08 8238 9044 - Thermo Fisher Scientific