Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-01092-1 |
Product Name/Description |
ichem VELOCITY Urine Chemistry Strips. An in vitro diagnostic medical device (IVD).
Reference Number: 800-7204
Lot Number: 7204089B
ARTG Number: 177551 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
9/11/2015 |
Responsible Entity |
|
Reason/Issue |
Beckman Coulter has become aware of a manufacturing defect where multiple strips with multiple missing pads were found inside the strip vial for the affected lot number. If the affected strips are used, it may result in quality control failures and potential false negative results for any or all analytes and erroneous pH results if affected strips are used to assay patient samples. |
Recall Action |
Recall |
Recall Action Instructions |
Beckman Coulter is advising users to discard affected lots. Any affected lots will be replaced by unaffected product. Beckman Coulter is advising users to consult with the Laboratory Director to determine if a retrospective review of results is clinically warranted. This action has been closed-out on 18/08/2016. |