TGA Recall Reference RC-2015-RN-01090-1
Product Name/Description Screw Inserter for TFN-Advanced (TFNA) Proximal Femoral Nailing System

Part number: 03.037.025

Lot Numbers: 8911887, 8911975, 9166356, 9285766, 9286671, 9319408, 9375245, 9375251, 9388359, 9388360, 9393071, 9393074, 9393075, 9393076, 9409795, 9409796, 9422892, 9422895, 9498455, 9310899, 9310900, 9319407

ARTG Number: 153950
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/11/2015
Responsible Entity Johnson & Johnson Medical Pty Ltd t/a DePuy Synthes
Reason/Issue Upon evaluation, DePuy Synthes discovered that certain lots of the Screw Inserter may potentially be etched with an incorrect graphic.

The orientation of the etched image of the oblique angle on the Screw Inserter may be reversed which does not correspond to the tip of the angle of the instrument. This could result in the oblique cut at the lateral end of the screw being at a 90 degree angle to the lateral cortex.

It is possible that the user would identify the final lateral oblique angle is not in the desired position, which could result in repositioning the screw within the nail so that the oblique angle of the screw is flush with the bone. The screw is designed to go in either orientation to allow proper depth placement of screw which is a critical factor for the success of the implant.
Recall Action Recall
Recall Action Instructions JJM is requesting users inspect all devices in their possession and to quarantine all non-conforming stock prior to their return to JJM. This action has been closed-out on 29/08/2016.
Contact Information 1300 562 711 - JJM Customer Service