Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01078-1
Product Name/Description Oxoid Anti-microbial Susceptibility Testing Disc CT0024B Gentamicin (CN 10) 5x50 discs

Lot Number: 1686823

ARTG Number: 208273
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 5/11/2015
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue Some discs in the affected lot may have insufficient concentration of antibiotic. This could result in an incorrect test reading due to small or absent zones of inhibition. Continued use of this lot could result in quality control failures, delayed results or false resistance reporting.
Recall Action Recall
Recall Action Instructions End users are requested to inspect their stock for affected units. Affected units are to be removed from further use. The requirement for the review of patient results should be determined by the laboratory director. This action has been closed-out on 06/06/2016.