Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01055-1
Product Name/Description Ultra-Fine Insulin Syringe for use with U-100 insulin

Catalogue Number: 326725

Lot Number: 5019750

ARTG Number: 125064
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 4/11/2015
Responsible Entity Becton Dickinson Pty Ltd
Reason/Issue The blister package for some insulin syringes may exhibit an open seal potentially impacting on product sterility.
No complaints or adverse events have been reported regarding this issue.
Recall Action Recall
Recall Action Instructions Users are requested to inspect stocks on hand and to quarantine affected units pending collection of the affected stock. A replacement or suitable credit will then be issued. This action has been closed-out on 31/08/2016.
Contact Information 1800 656 100 - BD Customer Service