Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-01049-1 |
Product Name/Description |
ADVIA Chemistry XPT Systems with software version 1.0.3. An in vitro diagnostic medical device (IVD)
Siemens Material Number (SMN): 10723034
ARTG Number: 175890 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
28/10/2015 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has identified multiple issues with the ADVIA Chemistry XPT systems with software version 1.0.3 that may affect the day to day behaviour and/or workflow of the system as follows:
Auto Start-Up Failing Calibration Interval Resets when a Reagent Blank (RBL) is run Control Definition screen assumes that the range is defined is 2 SD (Standard Deviation) Laboratory Information System (LIS) communication / Laboratory Automation (LAS) issue Printer Driver Resets ISE Calibration Ranges are too conservative for Urine Sodium (Na) Archiving and deletion may fail Workstation services may restart
With the alerts generated by the instrument for the scenarios impacting samples, the operator will be aware that a sample may not have been processed and can take action. Calibration and control issues are detectable through normal laboratory monitoring practices. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
All the issues above except (Issue: Workstation services may restart) are corrected in software version 1.1 which is now available for installation on affected systems. The temporary actions defined in the customer letter should be undertaken until the affected systems have been upgraded to software version 1.1. The remaining issue will be corrected in a future software version.
This issue has been corrected in software version 1.2.
This action has been closed out on 24/10/2018 |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |