Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-01044-1 |
Product Name/Description |
AU480, AU680, AU5800 Chemistry Analysers. An in vitro diagnostic medical device (IVD)
Part Numbers: MF7311 C cup
Various Model Numbers
All products manufactured since March 2011
ARTG Number: 236636 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
17/11/2015 |
Responsible Entity |
|
Reason/Issue |
The C cup may crack due to stress (e.g., wear during use), which may lead to coolant leaking from the cup. For CF150420-069 (customer complaint), the C cup cracked and leaked coolant onto the connectors of the PCB underneath the C cup. This resulted in a short-circuit of the PCB and smoke. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Beckman Coulter is inspecting all analysers and will replaced any affected cups. This action has been closed-out on 23/03/2018. |
Contact Information |
1800 060 881 - Customer Support Centre |