Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01044-1
Product Name/Description AU480, AU680, AU5800 Chemistry Analysers. An in vitro diagnostic medical device (IVD)

Part Numbers: MF7311 C cup

Various Model Numbers

All products manufactured since March 2011

ARTG Number: 236636
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 17/11/2015
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue The C cup may crack due to stress (e.g., wear during use), which may lead to coolant leaking from the cup.
For CF150420-069 (customer complaint), the C cup cracked and leaked coolant onto the connectors of the PCB underneath the C cup. This resulted in a short-circuit of the PCB and smoke.
Recall Action Recall for Product Correction
Recall Action Instructions Beckman Coulter is inspecting all analysers and will replaced any affected cups.
This action has been closed-out on 23/03/2018.