Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01025-1
Product Name/Description Roche Clinical Chemistry tests using NAD(H) or NADP(H) reaction. An in vitro diagnostic medical device (IVD).

Products Affected: CKMBL, Creatine Kinase-MB, GLDH3, Gen. 3, NH3 Unisys, Ammonia, NH3L, Ammonia

Material Numbers: 04525299190, 11929992216, 11877984216, 20766682322, 05975581190

Instruments Affected: COBAS INTEGRA, cobas c 501/502, MODULAR P, cobas c 701/702

ARTG Number: 174906
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 20/10/2015
Responsible Entity Roche Diagnostics Australia Pty Limited
Reason/Issue Temozolomide may cause interference in assays using NAD(H) or NADP(H). This has been confirmed in assays for GLDH, CKMB and NH3, depending on the analyser used. It has been determined that the interference occurs at either very low analyte, or too high drug concentrations. These internal investigations were carried out with the therapeutic peak plasma concentration of 0.013 mg/mL Temozolomide. Higher drug concentrations were not systematically tested, because no literature reference for plasma concentrations > 0.013 mg/mL was found.
The maximum bias for assays which showed interference at 0.013 mg/mL Temozolomide is shown below:

Parameter % bias
GLDH + 62%
CKMB - 31%
NH3 + 36%

This interference may results in falsely high or low results.
Recall Action Recall for Product Correction
Recall Action Instructions Roche is advising laboratories of the potential drug interference. Test instructions for use are in the process of being updated. This action has been closed-out on 19/08/2016.
Contact Information 02 9860 2357 - Roche Diagnostics