Type of Product Medical Device
TGA Recall Reference RC-2015-RN-01024-1
Product Name/Description Oxoid M.I.C Evaluator for Penicillin 256 microgram/ml

Catalogue Number: MA0101D/F

Lot Number: 1708082

ARTG Number: 208273
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 23/10/2015
Responsible Entity Thermo Fisher Scientific Australia & New Zealand
Reason/Issue A technical investigation has confirmed that the affected lot contains insufficient concentration of penicillin. Continued use of this product could result in the incorrect reporting of Minimum Inhibitory Concentration against penicillin.
Recall Action Recall
Recall Action Instructions Thermo Fisher is advising users to inspect stocks and remove from use any remaining units of the affected lot number. Affected stock can be returned or destroyed for credit.
Results issued with the affected lot number should be reviewed against case histories, other results and other relevant individual data. This action has been closed-out on 25/08/2016.
Contact Information 08 8238 9081 - Thermo Fisher Australia & New Zealand