Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-01000-1 |
Product Name/Description |
Zimmer Natural Nail CM Long- Cephalomedullary Nails
All product manufactured and packaged between July 2009 to July 2015
Multiple product codes and lot numbers affected
ARTG Number: 169894 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
13/10/2015 |
Responsible Entity |
|
Reason/Issue |
The Cephalomedullary Nails are packaged in double sterile barrier packaging that includes the placement of the product in a protective pouch and then into two sterile barrier trays (inner and outer). In approximately 5% of the tested samples, the outer tray was found to be compromised, however, there were no records which suggested that the sterility of the inner tray was affected. To date no complaints have been received regarding compromised packaging. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Zimmer is advising users to inspect current stock for damaged packaging and quarantine any affected units. Damaged product can be returned to Zimmer for replacement. This action has been closed-out on 22/08/2016. |
Contact Information |
02 9950 5426 - Zimmer |