Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00967-1
Product Name/Description Prism Medical P300 and C300 Lifts

Lifter P300
Models: 303050; 303051
Serial: P303470 – 479, P304158/164/171/172, P304536 – 539, P304866/867, P305116/118, P305125 – 127, P305130/134/135

Lifter C300
Models: 330050; 330053
Serial: C303221/224/225, CQ303107 – 116

Lifts manufactured prior to 8 June 2015

ARTG Number: 169639
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 30/09/2015
Responsible Entity Rehab Assist Pty Ltd
Reason/Issue Prism has been made aware of 17 incidents involving 300 series lifts in which the drive gear shaft has broken, eliminating the lifts ability to raise or lower a patient. Of these 17 incidents we have discovered 3 in which the free fall protection system did not engage properly. If both of these incidents occur at
the same time there is a risk for the person being supported by the lift to drop unexpectedly. The patients involved in the 3 incidents were not seriously injured and did not require medical attention.
Recall Action Recall
Recall Action Instructions Rehab Assist will replacing the affected units with a new unit. For further information please see https://www.tga.gov.au/alert/prism-medical-p300-and-c300-ceiling-hoist. This action has been closed-out on 18/08/2016.
Contact Information 03 9729 7978 - Rehab Assist