Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00961-1
Product Name/Description Alkaline Phosphatase (ALPAMP, ALPDEA and ALPA_c) used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems

Alkaline Phosphatase ALPAMP
Catalogue Number: 03035814
Lot Number: 327460

Alkaline Phosphatase ALPDEA
Catalogue Number: 03036535
Lot Number: 327465

Alkaline Phosphatase Concentrated ALPA_c
Catalogue Number: 06860450
Lot Number: 327471

ARTG Number: 176331
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/09/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue The reagent kit lots listed above demonstrate an increased incidence of absorbance flags (u, U) and //// errors. Absorbance flags u and U indicate abnormal high absorbance exceeding either the blank or sample limit. Absorbance flag //// indicates a calculation error and no result is generated by the system.
The increased rate of absorbance flags with these lots has not been associated with inaccurate results. If Calibration and Quality Control results are acceptable, reported patient results are acceptable. This risk to health is negligible.

Siemens advises the potential for clinical impact and overall risk to health as a result of this issue is negligible.
Recall Action Recall
Recall Action Instructions Customers are asked to quarantine affected units from use prior to destruction and/or disposal.
This action has been closed-out on 23/03/2018.
Contact Information 1800 310 300 - Siemens Technical Support Centre