Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00951-1 |
Product Name/Description |
Shuttle Discovery Seat (designed for children aged between six months and four years who have a physical disability)
Product Code: 2006-0010-000
Batch #: JS-06-14-DS-** JS-15-14-DS-** JS-18-14-DS-** JS-23-14-DS-**, JS-45-14-DS-** JS-46-14-DS-** ML-8-15-DS-** ML-13-15-DS-** ML-16-15-DS-** ML-17-15-DS-** ML-18-15-DS-** 61427 61738 ARTG #: 160111 |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
30/09/2015 |
Responsible Entity |
|
Reason/Issue |
Medifab have identified two safety issues relating to the Backrest and the Tilt in Space Locking Plunger on Shuttle Discovery.
1. On some Discovery Seats the Rear “S Tubes” are moving in and out excessively when the Backrest is rocked backwards and forwards. Although some movement is normal, there is potential for the S Tubes to come out of the Mounting Block should the motor vehicle used to transport the end user in the Discovery Buggy be involved in an accident.
2. On some Discovery Seats the Tilt in Space Plungers are not engaging in the Plastic Sides. There is potential that the Discovery seat may tilt back should the motor vehicle used to transport the end user in the Discovery Buggy be involved in an accident.
There have been no reported incidents in relation to these issues. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
A Technician or Product Consultant will contact customers to arrange a suitable time to carry out the corrective actions. For more details, please see https://www.tga.gov.au/alert/shuttle-discovery-seat . This action has been closed-out on 28/02/2018. |
Contact Information |
1300 543 343 - Medifab |