Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00874-1
Product Name/Description Flexi-Slip endotracheal tube stylets
(Intended for use in facilitating intubation of tracheal tubes.)

SAP/Catalogue Number: 502501-000060

Batch Numbers: 14BG27, 14FG03

ARTG Number: 214058
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 8/09/2015
Responsible Entity Teleflex Medical Australia Pty Ltd
Reason/Issue There have been reports of the plastic coating of the stylet splitting and/or breaking off of the stylet. This may result in a piece of plastic totally or partially occluding the patient’s airway and impairing ventilation, or necessitating invasive removal procedures in order to prevent complications such as atelectasis or pneumonia.

Over 200,000 units have been distributed and there have been 4 complaints to August 2015. This gives rise to a complaint rate of 0.00205%.

No patient injuries have been reported related to this issue.
Recall Action Recall
Recall Action Instructions Customers are asked to inspect their stock, quarantine any affected units from use and to follow the instructions on the acknowledgement form in order to have those units collected. An appropriate account credit will then be issued. This action has been closed-out on 22/08/2016.
Contact Information 1300 360 226 - Customer Service