TGA Recall Reference RC-2015-RN-00870-1
Product Name/Description EIZO Medical Monitors

Model Name: RX440, RX650, RX850 and GX540-CL

Affected Range:
RX440 - S/N 20001112 - 20931104
GX540 - S/N 20001013 - 25409114, S/N 20001283 - 20210324
RX650 - S/N 20001123 - 20766074
RX850 - S/N 20001024 - 20133074

ARTG Number: 115740
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 14/09/2015
Responsible Entity Arcadia Corporation Pty Ltd EIZO Oceania
Reason/Issue An internal quality control audit has revealed that image degradation (below 10bit) may occur when monitors are connected by display port on certain graphics boards. When starting up a computer or returning from sleep mode, an affected monitor displays 6-bit (64 graduations) instead of normal 8-bit (256 graduations) or 10-bit (1024 graduations).
This may not be recognised by the medical images, although it can be recognised by the grey-scale images and may affected diagnostic images. Monitors connected by DVI are not affected.
Recall Action Recall for Product Correction
Recall Action Instructions Eizo is providing interim instructions for users to follow to identify if the users are affected by the issue. A firmware update will be provided as a permanent fix. This action has been closed-out on 18/08/2016.