Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00840-1
Product Name/Description CA1:1L EVO 15 (Contrangle used in dentistry)

Part Number: 1600939-001

Serial Numbers: 15F0030, 15F0040, 15F0036

ARTG Number: 141205
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 7/09/2015
Responsible Entity William Green Pty Ltd
Reason/Issue There is the possibility of loosening of the dental tool, which may lead to potential safety risk if swallowed by the patient. The tool is not well held in the chuck system of the device.

This action has been undertaken prior to consultation with the TGA.
Recall Action Recall
Recall Action Instructions William Green is replacing all affected devices with unaffected product. This action has been closed-out on 11/04/2016.
Contact Information 1300 363 830 - William Green Pty Ltd