Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00818-1
Product Name/Description Cios Alpha mobile C-arm x-ray system

Catalogue/Lot number – 10308191

ARTG Number: 102180
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 28/08/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens have received reports that under certain circumstances, when starting the system up or reconnecting the C-arm cable to the main unit, the error 5000/243 “No X-ray - FD re-initialises” may occur. The error refers to a potential hardware malfunction of the Cios Alpha C-arm x-ray system that may occur when starting the system up or reconnecting the C-arm cable to the main unit. When this occurs, the system will become inoperable until the affected part is replaced.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is planning to release a modification for the Cios Alpha mobile C-arm x-ray system that will resolve the above-mentioned potential malfunction. As soon as the modification is available, a Siemens service engineer will contact customers to schedule implementation of the modification for your system.

Siemens is providing the following instructions to minimise the risk of this issue:
· Ensure that the system is ready after the startup
· Ensure that the error 5000/243 is not displayed, neither as message box nor in the FLC status bar (lower left side of the left monitor)
· In general, avoid disconnecting and reconnecting the C-arm cable during or shortly before starting a procedure. This action has been closed-out on 31/08/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre