Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00815-1
Product Name/Description WATCHMAN Left Atrial Appendage Closure Device with Delivery System
Material Numbers: M635WC21060, M635WC24060, M635WC27060, M635WC30060, M635WC33060, M635WS21060, M635WS24060, M635WS27060, M635WS30060, M635WS33060, M635WU21060, M635WU24060, M635WU27060, M635WU30060, M635WU33060

WATCHMAN Access System Access Sheath with Dilator
Material Numbers: M635TC10060, M635TC20060, M635TS10060, M635TS20060, M635TS40060, M635TU10060, M635TU20060, M635TU40060

ARTG Numbers: 216434 and 216435
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 27/08/2015
Responsible Entity Boston Scientific Pty Ltd
Reason/Issue BSC has identified that cross-threading of the haemostasis valve may occur if the valve is tightened with the dilator in place, potentially preventing subsequent sealing of the valve when desired.

This product correction reinforces existing Directions for Use (DFU) and provides further guidance regarding the correct use of the haemostasis valve in order to avoid cross-threading and to securely seal the valve, minimising the potential for undesirable blood leakage during surgery.
Recall Action Recall for Product Correction
Recall Action Instructions Boston Scientific is providing implanting surgeons with additional instruction to ensure that the risk of cross-threading of the haemostasis valve is mitigated. The product instruction for use are also being updated with this information.
Contact Information 02 8063 8100 - Boston Scientific