Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00767-1
Product Name/Description Alaris System Syringe Module

Model Number: 8110

Material Number: 10930269

ARTG Number: 175353
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 21/08/2015
Responsible Entity CareFusion Australia 316 Pty Ltd
Reason/Issue CareFusion has received service reports of a channel error on the Alaris System Syringe module model 8110.

A channel error is displayed on the syringe module in association with an audible and visual alarm on the attached PC Unit (error code 351.6740); once cleared on the PCU the Syringe module remains unresponsive to key presses. This error could occur during an infusion causing an interruption of infusion.

Whilst connected to an Alaris System PCU at the time of error, upon power down and reattachment to any other module the PCU will continue to function as expected, there is no fault or flow on effect to the PCU module.
Recall Action Recall for Product Correction
Recall Action Instructions CareFusion will adjust the syringe drive train assembly on the affected units and replace, if required. This action has been closed-out on 06/09/2016.
Contact Information 02 9624 9013 - CareFusion