Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00767-1 |
Product Name/Description |
Alaris System Syringe Module
Model Number: 8110
Material Number: 10930269 ARTG Number: 175353 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
21/08/2015 |
Responsible Entity |
|
Reason/Issue |
CareFusion has received service reports of a channel error on the Alaris System Syringe module model 8110.
A channel error is displayed on the syringe module in association with an audible and visual alarm on the attached PC Unit (error code 351.6740); once cleared on the PCU the Syringe module remains unresponsive to key presses. This error could occur during an infusion causing an interruption of infusion.
Whilst connected to an Alaris System PCU at the time of error, upon power down and reattachment to any other module the PCU will continue to function as expected, there is no fault or flow on effect to the PCU module. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
CareFusion will adjust the syringe drive train assembly on the affected units and replace, if required. This action has been closed-out on 06/09/2016. |
Contact Information |
02 9624 9013 - CareFusion |