Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00745-1
Product Name/Description Xia 3 Titanium Torque Wrench
Catalogue Number: 48237028
Lot Numbers: 098525,

Xia 3 Torque Wrench Short
Catalogue Number: 482397028
Lot Numbers: 11E035

ARTG Number: 160151
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/08/2015
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has received reports that the hex tip of the listed Torque Wrenches have been fracturing during final tightening. The hex tip of the wrench has broken off and separated from the main tube.

The manufacturer has identified four possible hazardous situations:
· Delay in surgery time (< 5 minutes) if an alternative torque wrench is needed
· Debris of non-implantable grade material falling into the wound
· Over/under-tightening of the blocker if an alternate instrument is not available
· Accidental implanting of the hex tip.
Recall Action Recall
Recall Action Instructions Customers are requested to quarantine all affected lots. A Stryker representative will contact all affected customers to arrange for collection of the affected products. Unaffected replacement units will be provided. This action has been closed-out 18/04/2017.
Contact Information 02 9467 1167 - RA & QA Co-ordinator