Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00745-1 |
Product Name/Description |
Xia 3 Titanium Torque Wrench Catalogue Number: 48237028 Lot Numbers: 098525,
Xia 3 Torque Wrench Short Catalogue Number: 482397028 Lot Numbers: 11E035
ARTG Number: 160151 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
17/08/2015 |
Responsible Entity |
|
Reason/Issue |
Stryker has received reports that the hex tip of the listed Torque Wrenches have been fracturing during final tightening. The hex tip of the wrench has broken off and separated from the main tube.
The manufacturer has identified four possible hazardous situations: · Delay in surgery time (< 5 minutes) if an alternative torque wrench is needed · Debris of non-implantable grade material falling into the wound · Over/under-tightening of the blocker if an alternate instrument is not available · Accidental implanting of the hex tip. |
Recall Action |
Recall |
Recall Action Instructions |
Customers are requested to quarantine all affected lots. A Stryker representative will contact all affected customers to arrange for collection of the affected products. Unaffected replacement units will be provided. This action has been closed-out 18/04/2017. |