Type of Product |
Medical Device |
TGA Recall Reference |
RC-2015-RN-00739-1 |
Product Name/Description |
ADVIA Chemistry System – Urea Nitrogen and Urea Nitrogen Concentrated Assays. An in vitro diagnostic medical device (IVD)
Urea Nitrogen Assay Catalogue Number: 03040257 Lot Numbers: 321973, 333872 Expiry date: 2015-11
Lot Number: 348126 Expiry date: 2016-06
Urea Nitrogen Concentrated Assay
Catalogue Number: 06860558
Lot Number: 321967 Expiry date: 2015-11
Lot Number: 332365 Expiry date: 2016-03
ARTG Number: 176642 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
12/08/2015 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has confirmed that Urea Nitrogen reagent kit lots listed and Urea Nitrogen concentrated reagent kit lots listed for use on ADVIA 1200, 1650, 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails. Failed calibration may lead to a slight delay in results until a replacement or suitable material to perform the calibration verification is obtained. The risk to health is negligible. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is advising users that if they experience this issue, to request unaffected replacement product. Siemens is not recommending a review of previously generated results. This action has been closed-out on 07/09/2016. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |