Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00739-1
Product Name/Description ADVIA Chemistry System – Urea Nitrogen and Urea Nitrogen Concentrated Assays. An in vitro diagnostic medical device (IVD)

Urea Nitrogen Assay
Catalogue Number: 03040257
Lot Numbers: 321973, 333872
Expiry date: 2015-11

Lot Number: 348126
Expiry date: 2016-06

Urea Nitrogen Concentrated Assay

Catalogue Number: 06860558

Lot Number: 321967
Expiry date: 2015-11

Lot Number: 332365
Expiry date: 2016-03

ARTG Number: 176642
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 12/08/2015
Responsible Entity Siemens Healthcare Pty Ltd
Reason/Issue Siemens Healthcare Diagnostics has confirmed that Urea Nitrogen reagent kit lots listed and Urea Nitrogen concentrated reagent kit lots listed for use on ADVIA 1200, 1650, 1800, 2400 and XPT Chemistry Systems demonstrate an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails. Failed calibration may lead to a slight delay in results until a replacement or suitable material to perform the calibration verification is obtained. The risk to health is negligible.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens is advising users that if they experience this issue, to request unaffected replacement product. Siemens is not recommending a review of previously generated results. This action has been closed-out on 07/09/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre