Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00724-1
Product Name/Description Access Thyroglobulin Reagent Kit. An in vitro diagnostic medical device (IVD)

For use with the Access Family of Immunoassay Systems including the Access 2, UniCel DxI 800 and UniCel DxI 600, UniCel DxC 600i, and the UniCel DxC 880i, UniCel DxC 860i, UniCel DxC 680i, and UniCel DxC 660i systems.

Product REF - 33860

Lot Number:
430969
431990
434727
522896
523251

Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/08/2015
Responsible Entity Beckman Coulter Australia Pty Ltd
Reason/Issue The Access Thyroglobulin reagent pack lots may generate falsely elevated results at concentrations <10 ng/mL. The estimated maximum magnitude of the potential error in the affected range is a shift of up to 0.36 ng/mL high at concentrations <10 ng/mL. Acceptable total imprecision was observed at values of 10 ng/mL and higher.
Recall Action Recall
Recall Action Instructions Beckman Coulter is requesting their customers to discard all Access Thyroglobulin reagent pack from the affected lots. Retesting should be carried out at the discretion of the laboratory director. This action has been closed-out on 26/08/2016.
Contact Information 1800 060 881 - Customer Support Centre