Type of Product Medical Device
TGA Recall Reference RC-2015-RN-00719-1
Product Name/Description Extension Set
(extension tubing for blood transfusion systems and standard intravenous administration sets)

REF 50.318H

Lot No. 20141226

ARTG Number 222156
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 13/08/2015
Responsible Entity BMDi TUTA Healthcare Pty Ltd
Reason/Issue BMDi TUTA Healthcare has been made aware of an issue with the female Luer lock (FLL) connector. When connected to the male luer lock (MLL) connector of another device the FLL may crack and subsequently leak. The reported incident rate is low and the fault has not been reproduced in house.
Recall Action Recall
Recall Action Instructions BMDi TUTA is requesting the customers to inspect, discontinue use, and quarantine affected stock for replacement. BMDi TUTA is arranging for the return of any remaining stock
Contact Information 02 9466 5300 - BMDi TUTA